3.4 Testing a mains cord
Visually inspect the mains cord and plug for
signs of damage.
Check that the correct fuse is fi tted.
If the cord passes a visual inspection proceed
with the electrical tests.
Plug the mains cord under test into the IEC
socket and the front panel test/mains socket
on the PrimeTest 250
+
.
Press the CLI test key.
The PrimeTest 250
+
will fi rst test the continuity
of the protective earth.
If the measured value is greater than the factory
set limit a cross is placed next to the R
PE
enunciator, a FAIL is indicated and the test
sequence will be halted.
If the measured value is less than the factory
set limit a tick is placed next the R
PE
enunciator.
The unit will proceed with the Insulation test.
If the Insulation Resistance is lower than the
factory set limit a cross is placed next to the
Riso enunciator and the test sequence is halted.
If the Insulation Resistance is greater than the
factory set limit a tick is placed next to the
Riso enunciator. If the Insulation Resistance is
less than the set limit a cross is placed next
to the Riso enunciator, the FAIL enunciator is
displayed and the test sequence is halted.
The unit will proceed with the wiring test,
checking the live and neutral conductors for
short or open circuit and reverse connections
If the wiring is correct a tick is placed next to
the cord enunciator, the GOOd enunciator
is illuminated and a PASS is indicated for
the sequence.