PrimeTest 100 Operating Instructions
17
4 Specification
Earth Continuity
Accuracy* ± (5% + 2 digits)
Test current 200mA minimum
Test voltage 9V nominal
Insulation resistance
Accuracy ± (5% + 2 digits)
Test voltage 500V
Test current >1mA into 500k
Test current <2mA into 2k
Leakage Current
Accuracy ± (5% + 2 digits)
Test voltage 40V rms, 50Hz AC
Test current <5mA into 2k
Cord Test
Earth continuity, insulation resistance as above.
Check for Live and Neutral open circuit, short
circuit or reversed polarity.
*When used with Seaward test lead, Part Number 347A002
Factory Set Pass/Fail limits
Class I Class II Cord
Earth
Continuity
0.2 ohms N/A 0.2 ohm
Insulation
Resistance
1.0Mohm 2.0Mohm
2.0Mohm
Leakage
0.75mA 0.25mA N/A